Line 72: Line 72:
<<Anchor(FoVGeometry)>>
Line 102: Line 101:
Further, for sources weaker than roughly Uranus the FF depends also on weather conditions, i.e. for the same source the derived FF is different for different sky loads. Below another comparison of the FF derived from a Uranus map observed during bad weather (#239 on Oct 20) with the same average FF as in the figures above. Further, for sources weaker than roughly Uranus the FF depends also on weather conditions, i.e. for the same source the derived FF is different for different sky loads. Below another comparison of the FF derived from an Uranus map observed during bad weather (#239 on Oct 20) with the same average FF as in the figures above.
Line 118: Line 117:
{{attachment:NIKA1mmPh6-20111018s55_indFFgfRef22_1_eqMxOverGfMx.png | beam deviation from Gaussian 1mm}} {{attachment:NIKA1mmPh6-20111018s55_indFFgfRef22_1_eqMxOverGfMx.png | beam deviation from Gaussian @1mm}}
Line 135: Line 134:

==== RZ, 26 Apr 2012 Response (fluxes) of the pixels ====

The two figures below show the peak and the integrated flux of all usable pixels in 9 Mars maps @1mm as function of the pixel number (1-116). Whereas the peak fluxes do not show any slope with the pixel number the integrated fluxes do show the same slope as their FWHM (see above Update Dec 9: beam broadening). This is not what one expect if the beam broadening was an optical effect: in this case the incoming power should be preserved, i.e. the peak fluxes should show a variation corresponding to the variation of the beam widths. Also important discussing effects of the optics: the deviations of the beam profiles from Gaussians (see above the figure 'beam deviation from Gaussian @1mm').

{{attachment:run3_1mm_2_gfHeigh.png | peak fluxes @1mm in 9 Mars maps}}
{{attachment:run3_1mm_2_gfInteg2.png | integrated fluxes @1mm in 9 Mars maps}}

In the coordinates of the FoV this slope shows the same picture as the beam broadening and the deviations of the beam profiles from Gaussians. Note further that this effect is not centered within the FoV.

{{attachment:compMbMxFF_1mmPh6_1_6.png | comparison of the flats derived from max and total fluxes}}

Left: flat-field derived using the peak fluxes, centre: flat-field derived using the total fluxes, right: the ratio of both.
Line 148: Line 160:

NIKA3 Offline Processing Results

Back to the NIKA3 main page

2011 October run

=Private results for the collaboration only=

Public results

List of offline processing tasks

Summary of Data reduction results

Here

FXD, 24 Oct 2011 - Photometric quality

Here are the maps obtained on one OTF scan with IRC 10216. Sky noise decorrelation is used.

IRC 10216 Scan65 18th 2mm map

IRC 10216 Scan65 18th 1mm map

The flux per kid is shown here using the calibration on Mars Scan #55. This is an indication that the new camera with its dIdQ modulation scheme provides a big photometric improvement with respect to 2010 campaign. The average flux is compatible with the range of fluxes measured on this (variable) source.

IRC 10216 Scan65 18th Flux per kid

RZ, 25 Nov 2011 - Pointing & focus effects

To derive the source positions I used the pointing maps available in the IRAM Multi Beam FITS (IMBFits) format. However, currently over 20% of the pointings are missing. The statistics shown below is therefore represantative but by far not complete.

gfPoiEq.png

Source positions in the eq offsets. Positions of planets are not shown. In red are shown positions of sources used for the pointing model, in black after the installation of the pointing model. The pointing run was apparently performed with Nasmyth offsets of EMIR ! The pointing offsets after the installation of the pointing model are still very large, even 3-5 times larger than the normal values.

gfPoiHor.png

As above in the hor system; positions of planets are also shown.

MarsUranus_azOff_extCorr.png MarsUranus_elOff_extCorr.png

Effekt of defocusing: the source moves by a couple of arcsec in azim & elev. Note the larger scatter of the elev-positions.

Both, the pointings and foci were done by mapping. However, to correctly process maps the correct beam parameters, i.e. the pixel positions and their flat field are mandatory. The effective FoV geometry, the flat field and their errors will come next.

Origin of the large pointing errors

All the errors and effects shown above and below (FoV geometry): scatter of the pixel positions, effect of the focus, imperfect pointing model a.s.o., cannot explain the very large (>10arcsec) pointing offsets. There must be still another problem I could not identify.

Update Feb 9 and 22:

To search for the source of the large pointing errors (and bimodal distribution in elev) I analysed also the pointings obtained with the heterodyne receivers. The poitings during the campaign are of mediocre quality but indicated some strange arcs in the plane (azCorrection,elCorrection). This finding is further confirmed by the (much clearer) examples below.

pointing_corr_EMIR.png

The pointing behaviour of the telescope may change rapidly, as shows the example below. This also happened during the NIKA run and explains the bimodal distribution of the pointing corrections.

change of pointing behaviour

Further informations gives the focus behaviour of the telescope (as the source position depends also on the focus, see above): the focus values show an unusual bimodal distribution around the mean value, which (in the example below) never appears as the true focus value. The difference between the mean focus values in the two branches of the distribution is >0.7mm !

focus behaviour

Important to note: such a pointing and focus behaviour of the telescope is new !

RZ, 30 Nov 2011 - Beam maps

To determine the pixel positions, the effective beam shapes and the flat field I used 9 Mars maps observed from Oct 17 to 22. Maps observed later on are of too bad quality. I could use sources weaker than Mars (Uranus, Neptune, ...) with sky noise filtering to determine the pixel positions & beams, but without the line-of-sight tau-corrections could not calculate the flat field. The pixels which show a clear cross-talk within the main beam were excluded.

FoV geometry

scatter of pixel positions average FoV 1mm

The figure above left (scatter of pixel positions) shows the derived pixel positions in each of the maps. Different markers and colors correspond to different maps. The scatter of the positions is larger in the centre and in the upper(left) part of NIKA, i.e this cannot be an atmospheric effect (anomalous refraction). For the few examples I checked, the changed positions were due to a cross-talk. The resulting average pixel positions are shown above right (fig. average FoV 1mm).

Centre of Nasmyth rotation

One very important point cannot be properly checked using the available data: the centre of the Nasmyth rotation. This centre is defined by the rotation axis of the mirror M3, but its determination depends also on the pointing behaviour. The ultimate proof would be if all the usable pixels would always show the source at one position, i.e. independent of the elevation and azimuth. For all coordinate transformations I had to assume this centre is close to pixel #21 (22 in my plots), as given by FXD.

Flat field

Update Dec 7: flat field for the beams in the near field.

Two flat fields are necessary for the data processing: for the far field (i.e. main beam = astronomical sources, MBFF) and for the near field (sky noise, i.e. the correlated signal, CSFF). The correct main beam flat should use integrals over the main beams. However, because the reason(s) for the beam broadening (see below) are not clear to me, I used just the peak fluxes, i.e. a flat field (PFFF) valid only for on-offs observed with not filled receiver arrays like MAMBO, LABOCA or SCUBA. Currently I also did not use any extinction corrections. The peak flux flat field shown below is therefore just indicative.

To calculate the flat for the beams in the near field in higher than first order approximation, just the slopes of the correlation distributions in data of weak sources are sufficient, i.e. when the assumption noise=data can be made (see "5 Dec 2011 - Instabilities, changed flat field, ..." below). Note that the calculation of the CSFF does not need any beam sizes.

Whereas each of the flat fields depends on the pixel response to the incoming signal, the ratio of the flat fields does not. In the no-problem case the ratio of the flat fields shows simply the change of the coupling of the receiver to the telescope. It may vary, e.g. due to a distortion of the FoV, but this variation cannot explain the observed ratios of >200%.

peakFlux flat field corrSig flat field

RZ, Feb 9 - Flat Field variations = calibration accuracy, i.e. stability of the system

The two figures below compare the beam parameters derived from the Mars map #54 and the Uranus map #149 (both on Oct 18) with the average beam parameters derived from 9 Mars maps. The upper inlet in both figures shows the difference in beam positions of each pixel (RPO = Receiver Pixel Offset). The differences of positions are within the possible accuracy. The flat field (FF) however shows very large discrepancy. Whereas the FF derived from the Mars map agrees within +-20% with the average FF, the FF derived from the Uranus map differs at the edges by a factor of up to 4.

comp_ff_Mars.png comp_ff_Uranus_goodSky.png

Further, for sources weaker than roughly Uranus the FF depends also on weather conditions, i.e. for the same source the derived FF is different for different sky loads. Below another comparison of the FF derived from an Uranus map observed during bad weather (#239 on Oct 20) with the same average FF as in the figures above.

comp_ff_Uranus_badSky.png

This demonstrates that the calibration per pixel varies by a factor of up to 4.

Beam broadening

Strange beam broadening is visible at 1mm (roughly a factor of 2) and much less at 2mm (below 50%). Two examples for Mars and Neptune are shown below.

beams 1mm Mars beams 1mm Neptune

Pixel #1 is not reliable. Note also the larger and elliptical beams of pixels #29 and 40 for Neptune. This is an effect of the cross-talk.

Towards the centre the beam profiles deviate stronger from a Gaussian than at the edges. This is most easily visible in a comparison of the maxima in the maps with the peaks of a Gaussian fitted to point-like sources.

beam deviation from Gaussian @1mm

The beam broadening is visible also in I & Q (below) and at 2mm (not shown).

beam broadening in I beam broadening in Q

For comparison, below are shown the beam sizes @1mm during run 2.

FoV and beams 1mm run2

Update Dec 9:

The beam broadening shows also a quite clear dependence on the pixel number. Below are shown the average FWHM of two first Mars maps @1mm.

FWHM(pixel number)

We can therefore ask: is the beam broadening a function of the pixel number, the FoV distribution therefore just a result of the choice of the pixel position, or vice versa ?

RZ, 26 Apr 2012 Response (fluxes) of the pixels

The two figures below show the peak and the integrated flux of all usable pixels in 9 Mars maps @1mm as function of the pixel number (1-116). Whereas the peak fluxes do not show any slope with the pixel number the integrated fluxes do show the same slope as their FWHM (see above Update Dec 9: beam broadening). This is not what one expect if the beam broadening was an optical effect: in this case the incoming power should be preserved, i.e. the peak fluxes should show a variation corresponding to the variation of the beam widths. Also important discussing effects of the optics: the deviations of the beam profiles from Gaussians (see above the figure 'beam deviation from Gaussian @1mm').

peak fluxes @1mm in 9 Mars maps integrated fluxes @1mm in 9 Mars maps

In the coordinates of the FoV this slope shows the same picture as the beam broadening and the deviations of the beam profiles from Gaussians. Note further that this effect is not centered within the FoV.

comparison of the flats derived from max and total fluxes

Left: flat-field derived using the peak fluxes, centre: flat-field derived using the total fluxes, right: the ratio of both.

Average beam sizes

For pointing sources weaker than roughly Neptune the errors of individual beams are too large to allow the check of the beam change across the FoV. Only the average beam shape can be fitted with sufficient accuracy. The figure below shows the major (triangles) and minor (circles) FWHM for all pointing sources available in the IMBFits format. There is a clear change of the observed beam size after the pointing run, as after the pointing run the sources appeared at different region of NIKA (recall: the pointing run was performed with the Nasmyth offsets for EMIR). This change is therefore an effect of the variation of the FWHM across the FoV. Within the errors there is no dependence of the FWHM on the source flux.

major and minor source sizes

Example beam maps obtained on Mars (~160Jy), Uranus (~33.5Jy) and Neptune (~12.6Jy) with NIKA1mm are show below. The contour levels are at 0.4, 0.8, 1.6 ... 25.6 and 50%, for Mars also at 0.1 and 0.2% (in yellow). The averaged FWHM is in all cases ~15arcsec.

MarsbeamMap 1mm Uranus beamMap 1mm Neptune beamMap 1mm

RZ, 1 Dec 2011 - The ''plateau''

Update Dec 7: figures @2mm

plateau@1mm uPixs Jy plateau@2mm uPixs Jy

Details of the plateau. The intensity scale is Jy/beam. The contour @1mm is at 9.5Jy/beam, @2mm 6.7y/beam.

plateau@1mm rows Jy plateau@1mm columns Jy

Intensity profiles @1mm per row (i.e. along the azimuth) and column (i.e. along the elevation) of pixels #22 (in red), #80 (in green), #102 (in cyan) and #105 (in blue).

plateau@2mm rows Jy plateau@2mm columns Jy

Intensity profiles @2mm per row (i.e. along the azimuth) and column (i.e. along the elevation) of pixels #11 (in red), #104 (in green), #105 (in cyan) and #91 (in blue).

plateau in I plateau in Q

The Mars map 20111019s52 in I and Q.

RZ, 5 Dec 2011 - Instabilities, changed flat field, nonlinearities

The correlation plots below show few examples of instabilities, changed flat field and nonlinearities of the 1mm signal

NIKA1mm-20111018s159_instab.png NIKA1mm-20111019s6_instab.png NIKA1mm-20111019s78_instab.png NIKA1mm-20111022s75_instab.png NIKA1mm-20111023s30_notLin.png

RZ, Feb 9 - Different apparent synchronization of 1mm and 2mm data

The 1mm data show a different apparent synchronization than the 2mm data. This is visible as a zig-zag of the source position from one subscan (i.e. a row in the map) to another. The two figures below demonstrate this. At 1mm the pixel #22, at 2mm the pixel #11 is shown. To remove the zig-zag visible in the 1mm data the time of the data has to be shifted by roughly -0.084sec, i.e. two samples. The 2mm data do not show this problem.

zig-zag@1mm.png no_zig-zag@2mm.png

RZ, update Mar 28: Different apparent synchronization of 1mm and 2mm data

The different apparent synchronization of the 1mm and 2mm data was caused by wrong times in the NIKA files and is visible in the R signals alone. The new files with corrected times do not show this discrepancy in the PF signals, i.e. both the 1mm and the 2mm show a zig-zag corresponding to roughly -0.084sec.

RZ, Feb 10 - Comparison of the R and PF signals

The calculated signals R and PF (which are supposed to represent the same incoming signal) differ considerably. Two examples - for a pixel in the centre (#10) and at the edge (#90) - @2mm are shown below. The ratio R/PF is not constant (i.e. the source shape is different in R & PF), it depends on the position within the FoV and is further band dependent (not shown).
The plots for the first NIKA FITS file with the R and PF signals provided by FXD were created with the FITS Viewer fv.

R for pixel#10@2mm PF for pixel#10@2mm R/PFvsMJD for pixel#10@2mm R/PFvsX for pixel#10@2mm R for pixel#90@2mm PF for pixel#90@2mm R/PFvsMJD for pixel#90@2mm R/PFvsX for pixel#90@2mm

In the centre the linearity of R(PF) breaks down at roughly 0.5 of the max signal, i.e. sources will appear broader in R than in PF. At the edge this effect is not visible. Qualitatively (but not quantitatively) a similar behaviour of R(PF) is visible also in the 1mm data (not shown).

R(PF) for pixel#10@2mm PF(PF) for pixel#90@2mm

This comparison shows there should be a beam broadening in the 2mm R data, but this is contrary to what is seen in the beam maps in R.

Obviously, in this example the R and the PF are simple inconsistent !

See below also "MC 30 Nov 2011 RF calculation on Mars", ....

RZ, Mar 22 - Comparison of R and PF: beam broadening and variation of the flat field across the FoV

The beam broadening and the variation of the Flat Field FF are clearly software effects. However, the question arises, is this *also* or *only* a software effect. I do not see how this could be answered for sure using just the astro data. Below I show the FWHM of the beams, the ratio of the beam angles and of the Flat Fields as derived using the R and PF signals for the first map with the PF signals provided by FXD; and at the bottom a standard comparison plot for another map.

2mm beams in R 2mm beams in PF

ratio of 2mm R & PF beam source angles ratio of 2mm R & PF flat fields

comparison of pixel parameters derived from R and PF

Comparison of receiver pixel parameters (RPPs) derived from R and PF. "IN" values were derived using PF, the values stored in the file were derived using R. FF = Flat Field, RPO = Receiver Pixel Offsets, beamAng = beam angle.

RZ, Mar 28 - Comparison of the R and PF signals: nonlineariry of PF(R)

The two figures below show the signal PF as function of R (PF(R), "phase 1" = R, "phase 2" = PF) for two consecutive scans on Oct 18. Whereas for scan 54 PF(R) is almost linear (R is slightly stronger than PF), for scan 55 PF(R) is highly nonlinear for most of the pixels. Note also the changed relative response of some pixels, i.e. the change of the Flat Field for these two maps.

PF(R)@2mm scan 54 Oct 18 PF(R)@2mm scan 55 Oct 18

Further, the difference PF-R is time variable. The two figures below show an on-on (more precisely position switch) observation. In the first figure is shown in red the R, in black the PF signal as a function of time. In the second figure is shown the difference PF-R. Note also the negative signals given by the pixels in the neighbourhood of #63.

on-off@2mm in R & PF on-off@2mm PF-R

RZ, update Apr 2

The figures below show the results of two on-on (more precisely position switch) observations, i.e. images in eq system. The numbers give the signal of Mars seen by the individual pixels. The negative signal (see also above) is an effect of the algorithm used to calculate the PF. Using the same Jy/counts calibration factor the on-ons give a factor of two lower peak flux for Mars than the maps. Note further the very bad shape of Mars, i.e. the "beam shape", slightly different in the positive and the negative signal. Such a bad beam shape is not visible in maps observed prior to these on-offs with the same focus setting. Both, the lower flux and the bad beam shape cannot be explained neither by focus effects nor by the rotation of the pixels across Mars. The reason for both problems is most probably also a software effect, i.e. an effect of calculation of the PF signal.

Mars_onOn_Oct18s60 Mars_onOn_Oct22s46 Mars_onOnPos_Oct18s60

SL, 07 Nov 2011 - Expected field distortion

grid distortion (updated Dec 1: 2 coefficients of the grid equation were switched)

FXD, 21 Nov 2011 - Technical note on kid frequencies

  • FXD, 24 Nov 2011 updated v2.1 (v2.1 Modulation explanation added, v2 Fits format exchange, v1 Basic frequencies) FXD, 24 Jan 2012 updated v2.2 (v2.2 Polynomial fit explained, fits header inputs from Albrecht)

Kid frequencies, fits format, polynomial fit

AM,MC 24 Nov 2011 Spectral transmissions

(IN/LPSC) Martin-Pupplet Interferometer Spectral Bandpass transmission

SL, 30 Nov 2011 - Calibration sources with PdB flux references

Calibrators fluxes for Oct 2010 (run #2) and Oct 2011 (run #3), fitted on PdB database (updated Dec 14, 2011)

MC 30 Nov 2011 RF calculation on Mars

RF calculation strategies

Blue line is the RFdIdQ as calculated in the standard way. Red, green and black are the results obtained using a 2-d polinomial fit of grade 2, 3 and 4 respectively.

MC 30 Nov 2011 dIdQ modulation on strong sources

dIdQ of modulation versus dIdQ of signal

The blue line is the trajectory on the IQ plane when observing mars (approx 7kHz signal). The red line is the observed DeltaI, DeltaQ, averaged on 100 points (please do not consider it. It doesn't have a lot of sense. I'll update.). Interesting are the green arrows, whiwh are the dI dQ obtained by the LO modulation: near the peak of the source, where we get far from the resonances, they get farther from the tangent of the IQ trajectory... to be investigated!

MC 07 Dec 2011 More comments on RF calculations

Amplitude IQ, pixel 5 Amplitude dIdQ, pixel 5 RF from CAMADIA vs RF from fit, pixel 5

Amplitude IQ, pixel 37 Amplitude dIdQ, pixel 37 [ATTACH]

The RF variation calculated by CAMADIA can differ from that obtained by a 2d fit used to find a relation F(I, Q). This in any case generally happens only for very strong signals (Mars, essentially...) and only for the pixels that happen to be already slightly distorted or in any case biased really to their limit. The images show 2 different pixels during a Mars observation. It can be seen that the one where there is a disagreement between RFdIdQ as evaluated by CAMADIA (blue line) and the one by the fit (red line) is the pixel with a distorted resonance (kid 5).

FXD 16 Dec 2011 Happy new year to the NIKA team : DR21 and DR21(OH) Preliminary data reduction (Sky noise decorrelation and then High-pass filtering by one subscan length)

DR21 at 1 and 2mm

DR21(OH) at 1 and 2mm

Nico 21 Dec 2011: Double resonnance separation + focal plane characterization

Here is a brief note on detectors decoupling and focal plane parameters reconstruction. Results are summarized in .fits files, one per configuration. I've only run the pipeline on two nights of observation for now but will update the list shortly. Comments on this preliminary work very much welcome. Merry christmas to you all !

FP_quicklook_111212.pdf d_2011_10_18_00h29m35_Uranus_333_FP_AB.fits d_2011_10_19_23h14m23_Uranus_220_FP_AB.fits

Nico 08 Feb 2012: update on focal plane characterization and distortions

Looking at 4 different OTF_Geometry scans throughout the calibration campaign, I compare focal plane reconstructions : they are very stable. I also look at the beam distortion pattern and show that it correlates very well to optical distortions. Comments welcome. Focal plane geometry fits files are available on the Neel Institute ftp website under the directory /Archeops/FPG.

FP_quicklook_120207.pdf

==== FXD 24 feb 2012 OnOffJitter Animation ======================

Animated data reduction of OnOff sequences

==== FXD, JMP 6 Mar 2012 Atmospheric Opacity calibration ======================

Analysis of frequency variations as a tool for the total power