Differences between revisions 17 and 18
Revision 17 as of 2009-01-26 13:33:20
Size: 873
Editor: mrt-aod
Comment:
Revision 18 as of 2009-01-27 01:25:37
Size: 889
Editor: visitor3
Comment:
Deletions are marked like this. Additions are marked like this.
Line 3: Line 3:
 * 2009-01-24, updated by DR, pointing from 14-01 to 24-01. attachment:pointing090124.png  * 2009-01-24, observer pointings between 14-01 to 24-01 (no model fit, DR) attachment:pointing090124.png

Heterodyne Pointing Model:

  • 2009-01-24, observer pointings between 14-01 to 24-01 (no model fit, DR) attachment:pointing090124.png
  • 2009-01-13, updated by JP, used 35 pointing scans with the B100 receiver. Only the pointing offset parameters P2 and P7 have been modified. The rms previous and after the fit was 1.9". The coverage and scatter of the pointing session is available in attachment:PointModel090113.pdf.

  • 2009-01-07, update by JP, 16 points, previous rms 9.5" and new rms = 3.4". attachment:point_model090108.pdf. This fit is confirmed by a fit by CK to 89 of in total 231 high-quality observers pointings which were conducted since beginning of this year: attachment:mrt-b100-2-080108.png
  • 2008-06-25, 29 points, previous rms = 2.58", new rms = 1.97". The coverage and scatter of the pointing session were: attachment:coverage.jpg attachment:scatter.jpg

PointingModelFits (last edited 2013-10-05 13:16:25 by mrt-lx7)