Back to the NIKA2 main page

NIKA2+ hardware adaptation to focuses differences

Historical context

In the first runs of NIKA2 in 2015 -ref-, we noticed that the 2mm beam was severely degraded and concluded that the dichroic #1 was severely distorted. After years using the dichroic #2, which was flat but had a problem of misplaced transition in the 1 mm band, and an unsuccessful try with the dichroic #3 which had a distortion problem similar to dichroic #1 -ref-, we finally came back to dichroic #1 for the NIKA2+ upgrade in March 2025 -ref-. This time the dichroic #1 was mounted differently than in the early runs 10 years before and than the dichroic #3 in 2017. The choice of mounting the dichroic #1 membrane in sandwich between 2 aluminum rings maintained in the copper mount with a copper flange screwed on the mount with a torque of 0.5 N*m, result from a series of tests conducted in IRAM lab these last years -ref-, which goal was to understand the cause of the distortion and find a way to suppress it or at least minimize it down to an acceptable value.
From the lab tests we were expecting a residual distortion in NIKA2 at cold temperature between 0 and 1 mm amplitude. The first order effect of the distortion is a displacement of the image surface in the cryostat, that can be compensated thanks to a displacement of the KID array. Beyond 1 mm distortion some second order aberration effect start to be more and more significant and degrade the beam even at the best focus.
A collection of copper rings between 1 and 10 mm thickness was fabricated, to be used as possible shims to be placed between the KID array holder and its mount on the cryostat, in order to shift the array position along the optical axis.

For the March 2025 intervention we changed a number of elements in the NIKA2 cryostat -ref-, but decided to not change the position of the arrays since we didn't knew exactly what would be residual distortion of dichroic #1 with its new mount in the NIKA2 cryostat environment.
During one year of using NIKA2 on sky, between run -ref- and run -ref-, we monitored the focus differences between the arrays. All these values are gathered in a document showing the evolution of the focus values and differences in a few plots -ref-. We noticed a constant focus offset between Array 1 and Array 3 (possibly due to a small residual distortion of the polarizer or a small difference in optical path length), and changing focus offset between Array 2 and the other arrays, passing from a small value with consequence on the beam to an unacceptable value where the 2 mm beam quality is very severely affected when the focus is set for the 1 mm Arrays. Somme Zemax simulations allows visualizing this effect with plots quantifying this effect -ref-. After each full warmup of the cryostat, the focus difference between Ar2 and the others at the following cool down was a bit more degraded, up to reaching an apparent asymptote at the last warmup in January 2026.

The residual distortion of the dichroic, its evolution with warmups and the possible measures to mitigate is were discussed at several NIKA2 meetings, in particular the following regular meetings -ref-, and the following dedicated one -ref-. In these meetings were discussed also another non optimal element in NIKA2+, the new Array 3, which is behaving better than the old one but performing significantly less good as the new Array 1, so that its contribution to the overall 1mm channel is weak. The discussions in the dedicated meeting leaded to create the following table of possible actions -ref-. From all these discussions we ended up with the following plan for this intervention in March 2026:

Actual plan for the intervention starting on March 10th, 2026

One week before the intervention, comparing focus values over the last year with typical EMIR focus values, Samuel came to the conclusion that the dichroic distortion must be upward and not downward as we thought. He came to this conclusion because in the optical design of NIKA2 the optimal focus for NIKA2 is obtained by increasing the distance between M1 and M2 by 1 mm compared to the optimal focus for EMIR. The value to enter in Zemax for NIKA2 to displace M2 in that direction is -1 mm, while in PaKo this is +1 mm. Several colleagues from IRAM Granada looked for information attesting with 100% certainty that a positive value entered in PaKo correspond to an increase of the distance M1-M2. They found some clues, like this drawing found by Miguel -ref-, but no document clearing totally the uncertainty.
A distortion upward of the dichroic shortens the optical path on the 2mm band, so that we should not introduce 10 mm thick shims to compensate to focus shift, but reduce the mount length by 10 mm, which is not possible. An alternative is to make the focus on Array 2 with M2, and insert shims on Ar1 and Ar3 to compensate for the corresponding focus shift. But as Zemax simulations show -ref-, this has consequences on the 1 mm beam quality. Hopefully, this affect the beam only by a few %, so that the loss is small enough to be acceptable. As a consequence the plan changed to the following actions:

Preliminary work before March 10th, 2026, and staff for the intervention

Intervention team: Alessandro Monfardini, Martino Calvo, Emile Prele (PhD student at Neel), David John, Samuel Leclercq

AM, MC, EP, DJ at the telescope on Monday, March 9th to prepare everything for the intervention on Tuesday morning.

March 10th, Tuesday

260310_00_Arrival_at_the_telescope_in_snowcat 260310_00_Telescope_living_room_obscured_by_snow 260311_31_Corner_of_living_room_window_emerge_from_snow

Measuring displacement of M2 with a laser, when changing focus Z value with OHB control system, and with PaKo:

260310_01_Laser_in_vertex_cabin_to_measure_distance_to_M2 260310_02_Laser_in_vertex_cabin_to_measure_distance_to_M2 260310_03_Laser_distance_to_M2_OHB_then_PaKo_focus-Z_0mm 260310_04_Laser_distance_to_M2_OHB_then_PaKo_focus-Z_+12mm

Laser in M1 vertex distance to M2 before any focus command: 13.061 m. - few mm in OHB control soft => 13.058 m ; 0 mm => 13.060 m ; +10 mm => 10.071 m ; -8 mm in PaKo => 13.052 m.
==> As expected, for both the OHB control software and in PaKo a positive value increases the M1-M2 distance. This means the dichroic distortion is upward, as feared.

Opening NIKA2:

260310_05_Start_opening_NIKA2 260310_06_Nose_removed_300K_body_flange 260310_07_Plug_for_nose_strut_behind_300K_body_flange 260310_08_Sliding_out_the_various_temperature_stages_shells 260310_09_Sliding_out_the_various_temperature_stages_shells 260310_11_Array2_on_its_7mm_thick_mount_disk 260310_12_2nd_Array3_about_to_be_dismounted

First SURPRISE: the dichroic is much more distorted than we expected, the distortion is visible with the naked eye !! As expected it is upward. We measure it:

260310_13_Measuring_dichroic_distortion_at_opening_Edge_16mm 260310_14_Measuring_dichroic_distortion_at_opening_Center_9mm

==> Distortion height = distance from a flat reference to the dichroic edge perpendicular to its surface - distance from a flat reference to the dichroic center perpendicular to its surface = 16-9 = 7 mm !!

This is logic that it's bigger at warm temperature where the membrane is at its maximal thermal expansion, than the 1.17 mm at cold temperature deduced from focus differences. But this is so spectacular that everyone here wonder whether we should really stick to the plan or rather reconsider it and try to do something about this distortion, namely dismounting the dichroic and reset its mechanical constraints to let it retrieve a flat shape. We decide to push a bit the Zemax simulations to weight various scenarios, and inform the persons involved in emails discussions since the discovery that the distortion should be upward and not downward the week preceding the intervention (NP, MS, SB, CK, AS, AB, MCF, IR, JT, JMP).

Other important observation looking at the Ar2 mount: it's possible to reduce its thickness by 2 mm, if we dismount it and mill it in the telescope machine shop. In terms of beam quality, it's preferable to do that rather than introducing thicker shims on the 1 mm arrays. We decide to go for it, Victor takes care of the milling.

After lot of discussion here we come with 2 options:

Option 2 is more uncertain result in terms of final Ar2 focus difference with the others and on its stability, but the potential gain in beam quality is bigger, which let more margins if we are a bit off. Option 1 has better certainty but not 100% (look at the fluctuations of recent Ar2 focuses), if it is successful the beam quality is lower than a successful option 2, and it gives no margin for being a bit off.

Several replying our email prefer option 1, we still hesitate at the telescope but have a small preference for option 2. Karl is put in the loop, calls us, and after explanations express also a preference for option 2. 3 arguments in additions to the one stated previously make the balance weight in favor of option 2, that we eventually decide to apply: [1] if we are a bit off, this is not so dramatic as only very few proposals require having both the 2mm beams and 1 mm beams optimal, for the other the focus can always be done on the most important band (we know this represents additional work for data processing, but the advantage is that the beam quality on the most important band would be better than with option 1)*, [2] if the dichroic continues to distorts more, we are not sure its physical integrity is insured, [3] we can learn much more for future optimization if needed with option 2) than with option 1).

* Update 1 week later after, done during the March 18th regular NIKA2 meeting: Stefano warns that such a solution is not possible, as it requires having 2 calibrations, with all it implies in terms of housekeeping and overheads, plus need to keep track of the focus configuration chosen for each scan. So if we are in a situation that is not optimal for all arrays, will have to decide on the best compromise focus for all arrays.

Dismounting dichroic to rest its distortion to 0. Second SURPRISE: after releasing the rings, the dichroic membrane does not recover a flat shape as it did in the lab !!

260310_16_Dichroic_distortion_after_loosening_its_rings_screws_50_to_0cN-m_Edge_16mm 260310_17_Dichroic_distortion_after_loosening_its_rings_screws_50_to_0cN-m_Center_13mm

==> Distortion height = distance from a flat reference to the dichroic edge perpendicular to its surface - distance from a flat reference to the dichroic center perpendicular to its surface = 16-13 = 3 mm! The membrane is not flat, but at least the distortion reduced by a factor 2.
After discussions on the various options offered to us we decide to fix again the membrane in its mount, applying a torque of 0.15 N*m torque on the screws of the copper flange instead of 0.5 N*m previously (the hope is that this gives more room to the dichroic to expand during warmup before being blocked again by the pressure of the rings, and thus reduce or even cancel the hysteresis effect that we suspect at each thermal cycling of the cryostat).

We decide to introduce a 3mm thick shims on Ar3, and not touch Ar1. So the total displacements between Ar3 and Ar2 is +3-(-2) = 5 mm. This is half the thickness of the thickest shim from the original plan. So if we managed to reduce the dichroic distortion by a factor 2, we should be right on spot in terms of focus equality between all arrays.

We mount the new array 3 with its 3mm shim, re-mount array 2 with its 2 mm thinned mount, and measure again the dichroic distortion after mounting it back in NIKA2.

260310_15_New_3rd_Array3_mounted_with_3mm_thick_shim 260310_18_Array2_mounted_back_on_its_mount_disk_2mm_thinner 260310_19_Dichroic_back_in_place_with_15cN-m_torque_on_rings_screws_Distortion_edge_16mm 260310_20_Dichroic_back_in_place_with_15cN-m_torque_on_rings_screws_Distortion_center_14mm 260310_21_Torque_wrench_used_for_dichroic_rings_screws

==> Distortion height = distance from a flat reference to the dichroic edge perpendicular to its surface - distance from a flat reference to the dichroic center perpendicular to its surface = 16-14 = 2 mm. It looks like the membrane relaxed a bit more the time to mount back the dichroic. Which is good.

End of Day 1.

260310_22_Sunset 260310_23_Dusk

March 11th, Wednesday

Closing NIKA2:

260311_25_Sliding_back_in_the_various_temperature_stages_shells 260311_26_Sliding_back_in_the_various_temperature_stages_shells 260311_27_Closing_NIKA2_cryostat_nose

Pumping, leak check, putting IKA2 back on its pinned position:

260311_28_Starting_pumping_NIKA2_cryostat 260311_29_Leak_checking_cryostat_vacuum 260311_30_Putting_NIKA2_on_its_pinned_position

Samuel refined Zemax simulations. The distortion of optical elements, like the dichroic has the first order consequence that leaded to the present intervention, of shifting the image surface along the optical axis. But it has several second order consequences which are harmless as long as the distortion is equivalent to a M2 focus shift lower than 1 mm, but that can be harmful to the data quality when the focus shift is higher than 1 mm. The most obvious, already presented in a document shown in the last regular NIKA2 meetings -ref- is an aberration of the beam such that even at the best focus the beam size and amplitude are not as sharp as it can be at best focus without the distortion. Another consequence not shown till now is that the distortion changes the size of the field of view (FOV) on the cold pupil and on the image surface. When the distortion is upward the FOV is smaller on the cold pupil and on the image, which means a larger illumination of the primary mirror (M1) at the risk of picking more spill over, and a reduction of the FOV on the KID array without change of the angular resolution (so the beams are bigger on the array). When the distortion is downward this is the opposite: smaller illumination of M1, less spill over, larger FOV on the image with smaller beams. Few screen shots to illustrate:

Upward distortion of the dichroic equivalent to 1.1 mm M2 focus shift, compensated by Ar2 displacement by -10 mm. Ray tracing layout on image surface and on cold pupil (contour of M1 for each field):

Dic_upward_dist_1p1mm_Image_at_-10mm_layout Dic_upward_dist_1p1mm_Cold_pup_layout

Downward distortion of the dichroic equivalent to 1.2 mm M2 focus shift, compensated by Ar2 displacement by +10 mm. Ray tracing layout on image surface and on cold pupil (contour of M1 for each field):

Dic_downward_dist_1p1mm_Image_at_+10mm_layout Dic_downward_dist_1p1mm_Cold_pup_layout

This shows that a downward distortion is more favorable. To keep in mind in case one day we want to consider trying to flip the dichroic now that it seems plastically distorted along one of its faces.

Now, if we are right on spot with the actions done during this intervention, that is to say distortion reduced by a factor 2, Ar2 position -2mm, Ar1 position unchanged, Ar3 position +3mm, all beams should be fine for the same M2 focus position; ray tracing layout on KID array, on cold pupil and spot diagram (with lowest Strehl ratio across FOV written on it) for Ar2, Ar1 and Ar3 respectively with M2 focus Z = 0.7 mm (e.i. -0.3 mm compared to the average best focus for Ar3 until now):

Dic_up_dist_0p6mm_dM2fZ-0p3mm_Ar2-2mm_layout Dic_up_dist_0p6mm_dM2fZ-0p3mm_2mm-band_Cold_pup_layout Dic_up_dist_0p6mm_dM2fZ-0p3mm_Ar2-2mm_Spot-Diagram
Dic_up_dist_0p6mm_dM2fZ-0p3mm_Ar1-0mm_layout Dic_up_dist_0p6mm_dM2fZ-0p3mm_1mm-band_Cold_pup_layout Dic_up_dist_0p6mm_dM2fZ-0p3mm_Ar1-0mm_Spot-Diagram
Dic_up_dist_0p6mm_dM2fZ-0p3mm_Ar3+3mm_layout Dic_up_dist_0p6mm_dM2fZ-0p3mm_1mm-band_Cold_pup_layout Dic_up_dist_0p6mm_dM2fZ-0p3mm_Ar3+3mm_Spot-Diagram

End of Day 2. No leak was detected, we started cooling down NIKA2.

260311_32_Pico_Veleta_with_snow 260311_33_30m_Telescope_with_snow 260311_34_Sunset

March 12th, Thursday

Cooling down is going well. Pressures seem a bit high after 12h compared to the usual after a full warmup, but this is probably due to out-gazing of the various internal components which were exposed to air during the opening of the cryostat.

Alessandro, Martino, Emile, Samuel leave the 30m. We'll know in a few days when NIKA2 is cold if all KID lines are OK (update March 16th: yes, all KID lines seem OK, box-S is out as expected, Box-P to watched). Then a new sweep will be necessary, at least on the new Array 3. Re-commissioning planned for the March 24-31 week -ref-.

New NIKA2+ configuration after intervention

Configuration after the intervention is as discussed in previous teleconfs and approved by IRAM. The actions were: a) replace array 3; b) move the focus of arrays 2 and 3; c) "reset" the dichroic shape (as best as possible). Here is the arrays configuration we've left inside the cryostat:

The status of the cryostat on March 13 is "cooling down", expected to be cold early next week (16/03/26). The priority will be to see if the resonances are OK on all the 19/20 lines where they are expected to be. In array 3 we expect if everything goes well, and considering that one line is missing, around 60-70% (700-800/1140) of the resonances.