Differences between revisions 10 and 12 (spanning 2 versions)
Revision 10 as of 2025-03-14 14:25:07
Size: 1564
Comment:
Revision 12 as of 2025-03-24 17:30:53
Size: 1694
Comment:
Deletions are marked like this. Additions are marked like this.
Line 12: Line 12:
Further info at the bottom of [[https://publicwiki.iram.es/DailyReportsNika2Upgrade2025|this site]]: March 14th sweep. Further info at the bottom of [[https://publicwiki.iram.es/DailyReportsNika2Upgrade2025|this site]]
Line 26: Line 26:

Array 1 (H, reflection) is NIKA2_1mm_MS_v3_10 - Al 25nm - LO 1.755 GHz
Array 3 (V, transmission) is NIKA2_1mm_MS_v3_05 - Al 22nm - LO 1.791 GHz

Back to the NIKA2 main page

Back to NIKA2 run 72 (Cryo run 84) main page

Noise traces right before the re-commissioning (March 14th sweep; arrays 1, 2 and 3)

noisetr_a1_20250314 noisetr_a2_20250314 noisetr_a3_20250314

Further info at the bottom of this site

NIKA2 transmission curves previously managed

- From preceding commissioning phases

- As given by Perotto et al. 2020:

NIKA2_transmission_2017_2024

- Transmission of the new arrays 1 and 3 according to lab measurements (provided by Alessandro Monfardini):

Lab_transmission_new_a1_2025 Lab_transmission_new_a3_2025

Array 1 (H, reflection) is NIKA2_1mm_MS_v3_10 - Al 25nm - LO 1.755 GHz Array 3 (V, transmission) is NIKA2_1mm_MS_v3_05 - Al 22nm - LO 1.791 GHz

InstrumentInformation (last edited 2025-03-24 17:35:30 by NikaBolometer)