This document describe the checks and test that '''must''' be done before and during the Bolometer pool, as well as the frequency needed for these tests. == MAMBO2/ABBA2 Checklist ==

This document describe the checks and test that '''must''' be done before and during the Bolometer pool, as well as the frequency needed for these tests. (Maintained by GQ.)
Line 9: Line 11:
 * Check that the paddle motor works.
Line 12: Line 16:
 * Teach the observers on the general aspects of the observation '''even''' they are expert MAMBO observers ('''weekly'''):  * Teach the observers on the general aspects of the observation '''even''' the expert MAMBO observers ('''weekly'''):
Line 19: Line 23:
 * Do absorber observations (AbsorberObs) when the weather is too bad even for heterodyne observations.
 * Do absorber observations (AbsorberObs) if the scans are too spiky
 * Observations on absorber ((AbsorberObs, '''at least once per week!''')
   * Do
when the weather is too bad even for heterodyne observations.
   * Do if the scans are too spiky
Line 25: Line 30:
 * Resistance  * Check the resistance
Line 27: Line 32:
 *
 * Check the Bias voltage
 * Check the blank/sync Signal at ABBA with an oscilloscope. ('''During tests, and/or if strong spikes appear''')
Line 40: Line 45:
    ||The skydip results are abnormally high||The paddle is put in from of the beam||Ask the operator to try it by hand or contact the Rx people||

MAMBO2/ABBA2 Checklist

This document describe the checks and test that must be done before and during the Bolometer pool, as well as the frequency needed for these tests. (Maintained by GQ.)

The tests are divided into different groups, as they supposed to be carried by different groups of the IRAM staff (operators, AoDs, or Rxs)

OPERATOR CHECKLIST

  • Bolometer temperature
  • Check that the paddle motor works.

AOD CHECKLIST

  • Teach the observers on the general aspects of the observation even the expert MAMBO observers (weekly):

    • CookbookPako.

    • When to switch to heterodyne
    • What is a spiky scan
    • Not to open a second monitor at flexible@mrt-lx3:monitor/
  • Do a backup of the DAFS (daily)

  • Observations on absorber ((AbsorberObs, at least once per week!)

    • Do when the weather is too bad even for heterodyne observations.
    • Do if the scans are too spiky

RXs CHECKLIST

  • Check the resistance
  • Check battery voltage
  • Check the Bias voltage
  • Check the blank/sync Signal at ABBA with an oscilloscope. (During tests, and/or if strong spikes appear)


PROBLEMS AND SOLUTIONS

This is a quick guide through normal problems that appeared in the past and the possible reasons and solutions.

  • PROBLEM

    REASON

    SOLUTION

    Bololog is not updated

    goPool has not been sent

    Ask the operator to do so

    Mopsic Monitor is not working

    There is other monitor running and the DAFS files have been corrupted

    Stop the second monitor and recreate the DAFS

    ABBA does not start getting data

    Wobbler is not wobbling

    Ask the operator

    ABBA does not start getting data

    Wobbler works fine but the signal did not reach ABBA

    Check the cable connections with ABBA

    ||

    The skydip results are abnormally high||The paddle is put in from of the beam||Ask the operator to try it by hand or contact the Rx people||