Line 25: Line 25:
   * On-off measurements on an absorber. The on-off are done w/o the wobbler taking offset in azimuth and elevation to activate the Basic or Cascade controller.
The Basic/Cascade limits are (600,400) arcsec in (Azimuth,Elevation). The Basic controller has a smoother behavior than the Cascade controller.
   * On-off measurements on an absorber. The on-off are done w/o the wobbler taking offset in azimuth and elevation to activate the Basic or Cascade controller. The Basic/Cascade limits are (600,400) arcsec in (Azimuth,Elevation). The Basic controller has a smoother behavior than the Cascade controller.

Tests performed during the heavy maintenance (2007.06.20)

TableOfContents

Goals

Identify the origin(s) of the "acceleration" problem in MAMBO2. We want to answer the following questions:

  • Is the acceleration problem due to a mechanical or electrical process ?
  • What is the influence of the driving program on its intensity (Basic/Cascade) ?

People

Pico Veleta: DJ, SL, SN, JP

Methods

  • Inclinometer installed on the Mambo2 platform

attachment:inclinometerPhoto.jpg


  • On-off measurements on an absorber. The on-off are done w/o the wobbler taking offset in azimuth and elevation to activate the Basic or Cascade controller. The Basic/Cascade limits are (600,400) arcsec in (Azimuth,Elevation). The Basic controller has a smoother behavior than the Cascade controller.

Results

  • Inclinometer measurements for two on-off measurements with the Cascade and the Basic controller.

BR attachment:inclinometerBolo.jpg

  • On-off measurements with the offsets in elevation: 360" (scan 47, Cascade) and 440" (scan 50, Basic)

BR Cascade controller attachment:scan47-elev-cascade.jpg


Basic controller attachment:scan50-elev-basic.jpg


  • An extra weight was put on the bolometer platform (scan 51)

attachment:scan51.jpg


  • Measurements of electrical peak with the oscilloscope during an on-off measurements with the offsets in elevation (scan 53).

No correlation was found between the beginning of the scans where the extra noise peaks appeared and the electric activity. A peak frequency of 5 sec was observed but which may be related to the air conditioned system.

Conclusions