Line 30: Line 30:
   * On-off measurements with the offsets in eleveation: 360" (scan 47, Cascade) and 440" (scan 50, Basic)    * On-off measurements with the offsets in elevation: 360" (scan 47, Cascade) and 440" (scan 50, Basic)
Line 42: Line 42:
---------
Line 43: Line 44:
   * Measurements of electrical peak with the oscilloscope during an on-off measurements with the offsets in elevation (scan 53).
No correlation was found between the beginning of the scans where the extra noise peaks appeared and the electric activity. A peak frequency of 5 sec
was observed but which may be related to the air conditioned system.
 

Tests performed during the heavy maintenance (2007.06.20)

TableOfContents

Goals

Identify the origin(s) of the "acceleration" problem in MAMBO2

People

Pico Veleta: DJ, SL, SN, JP

Methods

Inclinometer installed on the Mambo2 platform

attachment:inclinometerPhoto.jpg

Results

  • Inclinometer measurements for two on-off measurements with the Cascade and the Basic controller.

attachment:inclinometerBolo.jpg

  • On-off measurements with the offsets in elevation: 360" (scan 47, Cascade) and 440" (scan 50, Basic)

BR Cascade controller attachment:scan47-elev-cascade.jpg


Basic controller attachment:scan50-elev-basic.jpg


  • An extra weight was put on the bolometer platform (scan 51)

attachment:scan51.jpg


  • Measurements of electrical peak with the oscilloscope during an on-off measurements with the offsets in elevation (scan 53).

No correlation was found between the beginning of the scans where the extra noise peaks appeared and the electric activity. A peak frequency of 5 sec was observed but which may be related to the air conditioned system.

Conclusions