| Identify the origin(s) of the "acceleration" problem in MAMBO2 | Identify the origin(s) of the "acceleration" problem in MAMBO2. We want to answer the following questions: * Is the acceleration problem due to a mechanical or electrical process ? * What is the influence of the driving program on its intensity (Basic/Cascade) ?  | 
| Line 18: | Line 21: | 
| Inclinometer installed on the Mambo2 platform | * Inclinometer installed on the Mambo2 platform | 
| Line 21: | Line 24: | 
| ---------- * On-off measurements on an absorber. The on-off are done w/o the wobbler taking offset in azimuth and elevation to activate the Basic or Cascade controller. The Basic/Cascade limits are (600,400) arcsec in (Azimuth,Elevation). The Basic controller has a smoother behavior than the Cascade controller.  | 
|
| Line 27: | Line 32: | 
[[BR]]  | 
|
| Line 30: | Line 37: | 
| * On-off measurements with the offsets in eleveation: 360" (scan 47, Cascade) and 440" (scan 50, Basic) | * On-off measurements with the offsets in elevation: 360" (scan 47, Cascade) and 440" (scan 50, Basic) | 
| Line 32: | Line 39: | 
| [[BR]] | |
| Line 42: | Line 49: | 
| --------- | |
| Line 43: | Line 51: | 
|    * Measurements of electrical peak with the oscilloscope during an on-off measurements with the offsets in elevation (scan 53). No correlation was found between the beginning of the scans where the extra noise peaks appeared and the electric activity. A peak frequency of 5 sec was observed but which may be related to the air conditioned system.  | 
Tests performed during the heavy maintenance (2007.06.20)
Goals
Identify the origin(s) of the "acceleration" problem in MAMBO2. We want to answer the following questions:
- Is the acceleration problem due to a mechanical or electrical process ?
 - What is the influence of the driving program on its intensity (Basic/Cascade) ?
 
People
Pico Veleta: DJ, SL, SN, JP
Methods
- Inclinometer installed on the Mambo2 platform
 
attachment:inclinometerPhoto.jpg
- On-off measurements on an absorber. The on-off are done w/o the wobbler taking offset in azimuth and elevation to activate the Basic or Cascade controller.
 
The Basic/Cascade limits are (600,400) arcsec in (Azimuth,Elevation). The Basic controller has a smoother behavior than the Cascade controller.
Results
- Inclinometer measurements for two on-off measurements with the Cascade and the Basic controller.
 
BR attachment:inclinometerBolo.jpg
- On-off measurements with the offsets in elevation: 360" (scan 47, Cascade) and 440" (scan 50, Basic)
 
BR Cascade controller attachment:scan47-elev-cascade.jpg
Basic controller attachment:scan50-elev-basic.jpg
- An extra weight was put on the bolometer platform (scan 51)
 
attachment:scan51.jpg
- Measurements of electrical peak with the oscilloscope during an on-off measurements with the offsets in elevation (scan 53).
 
No correlation was found between the beginning of the scans where the extra noise peaks appeared and the electric activity. A peak frequency of 5 sec was observed but which may be related to the air conditioned system.