Line 5: Line 5:
== Preparation notes for the re-commissioning campaign (cryo run 92, NIKA2 80th; last update: 2026/03/20): ==
Line 6: Line 7:
= Preliminaries (last update: 2026/03/17): =  * A complete report about the intervention of NIKA2+ for hardware adaptation to focus differences can be read [[https://publicwiki.iram.es/NIKA2%2BFocDiffAdaptWork|here]]. In principle, array 1 was not touched during the NIKA2+ intervention. Similar focus corrections are expected for all three arrays.
Line 8: Line 9:
 * According to DJ, a new sweep is provided before the beginning of this re-commissioning week. We will proceed as usual: (i) point/focus with the old sweep (run 91), (ii) get beam maps with the new sweep (run 92), (iii) then observe with the old sweep until the new DAFs/Kidpars are ready (it usually takes few hours).  * A new sweep, only for Ar3, was performed on 2026/03/19; see the final part of the previous link. Eventually (see Note below), we would proceed as usual: (i) point/focus with the old sweep (cryo run 91), (ii) get beam maps with the new sweep (cryo run 92), (iii) then observe with the old sweep until the new DAFs/Kidpars are ready (it usually takes few hours).
Line 12: Line 13:
 * A correction of parameters P2 and P7 of the pointing model of NIKA2 is provided on Monday 23/3/2026 (based on the round mean of the most recent pointing corrections: ΔAz. = 4", ΔEl. = 6")

 * A complete report about the intervention of NIKA2+ for hardware adaptation to focus differences can be read [[https://publicwiki.iram.es/NIKA2%2BFocDiffAdaptWork|here]]. In principle, array 1 was not touched during the NIKA2+ intervention. Similar focus corrections are expected for all three arrays.

= Notes for the next (run 80) re-commissioning campaign (under construction): =
 * A correction of parameters P2 and P7 of the pointing model of NIKA2 is provided on Monday 23/3/2026 (based on the round mean of the most recent pointing corrections: ΔAz. = 4", ΔEl. = 6"). '''Update:''' done.
Line 20: Line 17:
 * If the opportunity arises/a decision is made to continue observations of the NIKA2 pool, or even heterodyne pool programs, use the accounts nikaw-25 or poolw-25, respectively.  * If the opportunity arises, observations from the NIKA2 pool could be continued, or even heterodyne pool programs if conditions are not favorable for useful observations with NIKA2. In both cases, the instructions for the current projects have been updated.
 
== PIIC DAFs status ==

 * DAFs for cryo run 91 were installed on 2026/03/19 by Stefano Berta as default for PIIC at Pico Veleta. When setting up the PIIC environmental variables with ''gagpiic'' and starting PIIC, the environment variable $MOPSIC_DAFS points by default to /sw/gildas/mopsic/dafs.
Line 24: Line 25:
 * New sweep (run 92), likely only for Ar3 (19/3/2026 update: ready; see the last record [[https://publicwiki.iram.es/NIKA2%2BFocDiffAdaptWork|here]].
 * Preliminary check of the system on the sky, getting first pointing/focus corrections
 * Produce high-quality DAFs/RPPs and Kidpars
 * Determine the optimal Z focus position and study the new focus surface per array
 * Produce new final NEFDs, in total power and polarization (as possible)
 * Preliminary verification of the system in the sky, obtaining the first pointing/focus corrections and quantifying the first focus deviations (hopefully small) between pairs of arrays.
 * Determination of the optimal Z-axis focus position and study of the new focus surface for each array.
 * Produce high-quality DAF/RPP and Kidpars.
 * Produce new final NEFDs, in total power and polarization (if possible), and providing an estimate of the fraction of valid KIDs per array. Among other aspects, both are important for updating the NIKA2 Time Estimator.
Line 30: Line 30:
== Preparation notes == == Observation priorities ==
Line 32: Line 32:
== Schedule ==

== Staff ==
 * Obtaining high-quality beam maps (at least 2) in total power/polarimetry.
 * Getting calib_1scan maps on Uranus, Mars and secondary flux standards/QSOs (at different elevations, day/night) in total power/polarimetry + skydips.
 * Once a couple of good beam maps on focus are obtained, getting de-focused beam maps, if possible: 5 focus positions, at least one beam map for each focus position.
 * Getting a small deep field centered on a moderately bright source (few hours, say at least 5) in total power and polarimetry.
 * Continuing with science observations corresponding to the pools in force, depending on weather conditions.

Back to the NIKA2 Run 80 main page

Preparation notes for the re-commissioning campaign (cryo run 92, NIKA2 80th; last update: 2026/03/20):

  • A complete report about the intervention of NIKA2+ for hardware adaptation to focus differences can be read here. In principle, array 1 was not touched during the NIKA2+ intervention. Similar focus corrections are expected for all three arrays.

  • A new sweep, only for Ar3, was performed on 2026/03/19; see the final part of the previous link. Eventually (see Note below), we would proceed as usual: (i) point/focus with the old sweep (cryo run 91), (ii) get beam maps with the new sweep (cryo run 92), (iii) then observe with the old sweep until the new DAFs/Kidpars are ready (it usually takes few hours).
  • Note: if the new sweep is only for Ar3, we won't need to switch between new and old sweep, because Ar2 and Ar1 will be available for pointing and focus :-)

  • A correction of parameters P2 and P7 of the pointing model of NIKA2 is provided on Monday 23/3/2026 (based on the round mean of the most recent pointing corrections: ΔAz. = 4", ΔEl. = 6"). Update: done.

  • For re-commissioning tests/observations, NIKA2 account t22 should be used (~/pako/202503/).
  • If the opportunity arises, observations from the NIKA2 pool could be continued, or even heterodyne pool programs if conditions are not favorable for useful observations with NIKA2. In both cases, the instructions for the current projects have been updated.

PIIC DAFs status

  • DAFs for cryo run 91 were installed on 2026/03/19 by Stefano Berta as default for PIIC at Pico Veleta. When setting up the PIIC environmental variables with gagpiic and starting PIIC, the environment variable $MOPSIC_DAFS points by default to /sw/gildas/mopsic/dafs.

Main Goals

  • Preliminary verification of the system in the sky, obtaining the first pointing/focus corrections and quantifying the first focus deviations (hopefully small) between pairs of arrays.
  • Determination of the optimal Z-axis focus position and study of the new focus surface for each array.
  • Produce high-quality DAF/RPP and Kidpars.
  • Produce new final NEFDs, in total power and polarization (if possible), and providing an estimate of the fraction of valid KIDs per array. Among other aspects, both are important for updating the NIKA2 Time Estimator.

Observation priorities

  • Obtaining high-quality beam maps (at least 2) in total power/polarimetry.
  • Getting calib_1scan maps on Uranus, Mars and secondary flux standards/QSOs (at different elevations, day/night) in total power/polarimetry + skydips.
  • Once a couple of good beam maps on focus are obtained, getting de-focused beam maps, if possible: 5 focus positions, at least one beam map for each focus position.
  • Getting a small deep field centered on a moderately bright source (few hours, say at least 5) in total power and polarimetry.
  • Continuing with science observations corresponding to the pools in force, depending on weather conditions.

GoalsPrepStaffNika2Run80 (last edited 2026-04-03 12:59:22 by NikaBolometer)