NIKA2 Re-Commissioning (3rd. attempt)
NIKA2 run 74, Cryo run 86
Infos about the PIIC DAFs for the PIIC monitor
According to the results obtained from processing data of June 2025 NIKA2 re-commissioning given here, DAFs RPP scheme used during run 73 is kept. Therefore, two different sets of DAFs for Ar1+3 and for Ar2 are used. In practice:
- Ar2
- open a terminal, go to the QL directory
- gagpiic
- export MOPSIC_DAFS=/local/users/flexible/reduce/dafs_bck/dafs_local_tmp_run73_ar2
- start the Ar2 monitor (pf)
- repeat in another terminal for the Ar2 monitor(m)
- Ar1 and Ar3
- open a new terminal, go to the QL directory
- gagpiic
echo $MOPSIC_DAFS -> check that it says /sw/gildas/mopsic/dafs
- if not, then set it to that directory
- proceed normally with the Ar1 and Ar3 monitors
July 12th, 2025
These tests are being carried out in the framework of the first Heterodyne pool week of Summer 25. Thanks to all involved for their comprehension.
Albrecht (AoD) and Angel at the telescope. Operators on Duty: Gloria (night) and Kike. Cryostat ok (T_MC = 0.170 K, T_Still = 1.21 K, T_4K = 8.37 K).
DAQ started with the window closed for sampling some instrumental noise traces. Bad KIDs (Ar1/Ar2/Ar3): 12, 5, 14. Screenshots of noise traces, in that order: