|| {{attachment:oldconfrun8.png||width=800}} || {{attachment:newconfrun8.png||width=800}} || || {{attachment:oldconfrun8.png||width=700}} || {{attachment:newconfrun8.png||width=700}} ||
Line 13: Line 13:
'' '''Figure.''' Left: old config, right: new config. MADE WITH WINDOW ON CRYOSTAT, TO BE CONFIRMED (see tomorrow) WITH LN2. '''DON'T USE THIS BRAIN!!''' '' '' '''Figure.''' Left: old config (2mm), right: new config (2mm). MADE WITH WINDOW ON CRYOSTAT, TO BE CONFIRMED (see tomorrow) WITH LN2. '''DON'T USE THIS BRAIN!!''' ''
Line 15: Line 15:
Strong (strange) 12Hz noise on some (e.g. 4-5) pixels. Strong (strange) 12Hz noise on some (e.g. 4-5) pixels. See tomorrow daily report for more on that.
Line 21: Line 21:
8h00. Theoretical start of the run. Winds peaking at 40m/s, telescope stopped probably the whole morning. The ski station is closed due to the wind. '''8h00.''' Theoretical start of the run. Winds peaking at 40m/s, telescope stopped probably the whole morning. The ski station is closed due to the wind.
Line 23: Line 23:
9h00. LN2 in front of the cryostat window to play at least with the resonances. "Decalage" found by the new "restore tones" is good. -22kHz and -1KHz at 1 and 2mm respectively, indicating that the param file run8.txt is well representative of LN2 temperature. '''9h00.''' LN2 in front of the cryostat window to play at least with the resonances. "Decalage" found by the new "restore tones" is good. -22kHz and -1KHz at 1 and 2mm respectively, indicating that the param file run8.txt is well representative of LN2 temperature.
Line 31: Line 31:
|| {{attachment:run8LN2_1mm.png||width=800}} || {{attachment:run8LN2_2mm.png||width=800}} || {{attachment:run8LN2_1mm.png||width=1200}}
Line 33: Line 33:
'' '''Figure.''' Left: 1mm new config, right: 2mm new config. MADE WITH LN2 close to pupil, TO BE CONFIRMED ON THE SKY. When we can get it. '' {{attachment:run8LN2_2mm.png||width=1200}}
Line 35: Line 35:
13h00. Wind hits 42m/s (150km/h). The telescope basement is a block of ice. Put again the window on the cryostat, "restore tones" works very well. Found a shift of -393kHz and 100kHz. '' '''Figure.''' Top: 1mm new config (LN2). Bottom: 2mm new config (LN2). MADE WITH LN2 close to pupil, TO BE CONFIRMED ON THE SKY. When we can get it. ''
Line 37: Line 37:
15h. Went to the cabin to try understanding the monochromatic lines (e.g. 8Hz, 12Hz) on some pixels. Difficult to conclude since it apperars that '''even with the window closed''' (but without Al tape) you see large signals due to reflections of body temperature through the HDPE then onto the curved inner surface of the window and into the cryostat. '''13h00.''' Wind hits 42m/s (150km/h). The telescope basement is a block of ice. Put again the window on the cryostat, "restore tones" works very well. Found a shift of -393kHz and 100kHz.

'''15h00.'''
Went to the cabin to try understanding the monochromatic lines (e.g. 8Hz, 12Hz) on some pixels. Difficult to conclude since it apperars that '''even with the window closed''' (but without Al tape) you see large signals due to reflections of body temperature through the HDPE then onto the curved inner surface of the window and into the cryostat.

Friday 14th February 2014

Xavier and Alessandro arrive at the telescope. Made a new parameter file FOR THE 2MM array ONLY (definitely need the sky for the 1mm).

Old configuration (2mm): DAC gains: 20-23-20-28-35. General niveau: 5.

New configuration (2mm): DAC gains: 20-23-20-38-55. General niveau: 5. Pixels 17-25: niv 6. Pixels 55 to 56: niv 6. Pixels 73 to 78: niv 4. Pixels 96 to 108: niv 6.

oldconfrun8.png

newconfrun8.png

Figure. Left: old config (2mm), right: new config (2mm). MADE WITH WINDOW ON CRYOSTAT, TO BE CONFIRMED (see tomorrow) WITH LN2. DON'T USE THIS BRAIN!!

Strong (strange) 12Hz noise on some (e.g. 4-5) pixels. See tomorrow daily report for more on that.

Saturday 15th February 2014

8h00. Theoretical start of the run. Winds peaking at 40m/s, telescope stopped probably the whole morning. The ski station is closed due to the wind.

9h00. LN2 in front of the cryostat window to play at least with the resonances. "Decalage" found by the new "restore tones" is good. -22kHz and -1KHz at 1 and 2mm respectively, indicating that the param file run8.txt is well representative of LN2 temperature.

NEW PARAM FILE MADE ON LN2

LN2 configuration (1mm): DAC gains: 20-23-20-30-40. General niveau: 6. Individual levels to be adjusted on the sky but it'll be difficult. Bad pixels: too difficult, see offline data analysis.

LN2 configuration (2mm): DAC gains: 20-23-20-33-50. General niveau: 5. Pixels KB021,30,41,42,111 at niv. 6. Pixels KB073-80,83,88,111,114-116,119-120 at niv.4. Bad pixels: KB033,75,76,88,109,110.

run8LN2_1mm.png

run8LN2_2mm.png

Figure. Top: 1mm new config (LN2). Bottom: 2mm new config (LN2). MADE WITH LN2 close to pupil, TO BE CONFIRMED ON THE SKY. When we can get it.

13h00. Wind hits 42m/s (150km/h). The telescope basement is a block of ice. Put again the window on the cryostat, "restore tones" works very well. Found a shift of -393kHz and 100kHz.

15h00. Went to the cabin to try understanding the monochromatic lines (e.g. 8Hz, 12Hz) on some pixels. Difficult to conclude since it apperars that even with the window closed (but without Al tape) you see large signals due to reflections of body temperature through the HDPE then onto the curved inner surface of the window and into the cryostat.