Differences between revisions 11 and 12
Revision 11 as of 2014-02-15 08:12:03
Size: 1056
Comment:
Revision 12 as of 2014-02-15 11:21:39
Size: 1158
Comment:
Deletions are marked like this. Additions are marked like this.
Line 13: Line 13:
'' '''Figure.''' Left: old config, right: new config.'' '' '''Figure.''' Left: old config, right: new config. MADE WITH WINDOW ON CRYOSTAT, TO BE CONFIRMED (see tomorrow) WITH LN2. '''DON'T USE THIS BRAIN!!''' ''

Friday 14th February 2014

Xavier and Alessandro arrive at the telescope. Made a new parameter file FOR THE 2MM array ONLY (definitely need the sky for the 1mm).

Old configuration (2mm): DAC gains: 20-23-20-28-35. General niveau: 5.

New configuration (2mm): DAC gains: 20-23-20-38-55. General niveau: 5. Pixels 17-25: niv 6. Pixels 55 to 56: niv 6. Pixels 73 to 78: niv 4. Pixels 96 to 108: niv 6.

oldconfrun8.png

newconfrun8.png

Figure. Left: old config, right: new config. MADE WITH WINDOW ON CRYOSTAT, TO BE CONFIRMED (see tomorrow) WITH LN2. DON'T USE THIS BRAIN!!

Strong (strange) 12Hz noise on some (e.g. 4-5) pixels.

Saturday 15th February 2014

8h00. Theoretical start of the run. Winds peaking at 40m/s, telescope stopped probably the whole morning.

9h. LN2 in front of the cryostat window to play at least with the resonances. "Decalage" found by the new "restore tones" is good. -22kHz and -1KHz at 1 and 2mm respectively, indicating that the param file run8.txt is well representative of LN2 temperature.

DailyReports8 (last edited 2014-10-02 15:54:44 by NikaBolometer)